Picture Name

Products

+
  • e2s-45.jpg

Oil rotary pump E2S45

The E2S series (E2S45, E2S65 and E2S85) vane pumps provide an ideal balance of performance, cost and availability. Reliable design, simple structure, especially suitable for a variety of standard applications requiring medium vacuum. It can quickly evacuate and process steam, helping to reduce noise levels in the workplace to a greater extent.

Classification:

E2S two-stage rotary vane pump E2S

two-stage oil rotary vane pump

Key words:

E2S two-stage rotary vane pump E2S


Contact Us
  • Product features and uses
  • Product parameters and graphs
  • | Features

    1, Professional Performance:

    High Pump Speed Shortens Cycle Time and Improves Productivity

    Hydro Mode Improves Steam Handling Capacity to Suit Wet Application Environment

    Low Noise, Meeting Laboratory Environment Requirements

    2, Simple Modern Design:

    Economical product with simple layout

    Built-in oil filter can prevent pollution and oil leakage of the pump mechanism

    Innovative design can improve cooling performance and maintain excellent steam resistance

    No complicated mechanical arrangement can keep the technology simple

    3, excellent usability:

    Effective product design to meet user needs

    Simple and convenient to use and maintain. It can be cleaned on site to maximize normal operation time.

    Efficient overheat protection and low oil level detection (optional function)

    Avoid operator misoperation and higher safety.

    4, E2M and EH booster pump combination:

    E2S two-stage rotary pump and EH supercharger standard combination can meet high capacity requirements. These pumping systems are pre-assembled at the factory and are compact, mobile and easy to use. The E2S85-EH1200 can be started directly at atmospheric pressure, providing an effective speed of 900 m 3 /h (@ 60Hz), and can reach a high vacuum range, the ultimate vacuum is 2x 10 -4 mbar.

    | Use

    Use in a variety of standard applications that require a medium vacuum

  • | specification parameter

    type E2S45 E2S65 E2S85
    extraction speed (m /h) 50Hz 42 57 78
    60Hz 50 69 94
    displacement upper limit value (m/h) 50Hz 47 64 87
    60Hz 56 77 105
    limit pressure (Pa)(no gas town) 50Hz ≦ 0.3
    60Hz
    limit negative pressure 50Hz 0.6
    60Hz with gas town
    Maximum Allowable Outlet Pressure bar (g) 0.5
    Motor Voltage V 100-120V/200-230V, Single Phase, 50/60Hz, 100-120V/200-230V, single phase, 50/60Hz, Single Phase, 50/60Hz, AC100-230V ±10%
    3 Phase, 50/60Hz, 200-230V/380-460V
    Rated Power (50/60Hz) kW 1.1/1.3 1.5/1.8 2.2/2.6
    Water Vapor Tolerance Upper Limit Value (Kg/h) 50Hz 0.62 0.66 0.73
    60Hz 0.74 0.8 0.87
    oil (upper limit) L 4.3 4.8 5.5
    Motor Protection Level IP 55
    Noise dB(A) 50Hz 58 60 60
    60Hz 60 62 62
    Inlet ISO-F/NW 40 40 40
    Exhaust mm 25 25 25
    Weight kg 80 90 100
    Remarks Recommended Oil Ultragrade Performance 70

    | Curve

Related Products


Immediately consult