  • Qulee HGM2_202.png

Gas Analyzer HGM2-202

HGM2 series analyzer belongs to high-end models, suitable for various research and development purposes.

Classification:

Aico

gas analyzer

High performance process

gas analyzer

Key words:

Aico


Contact Us
  • Product features and uses
  • Product parameters and graphs
  • | Features

    1, Measurement Quality Range 1~200 amu
    2, Ideal Choice for Various Research and Development Applications-Realize Highest Sensitivity 2.5 × 10-6A/Pa
    3 as High-end Function, Communication Interface-Support Ethernet Communication
    4, Integrated Design of Display Part Can Measure
    5 without Computer, Simple Operation-One Click Function (It's easy for everyone, no complicated operation is required)
    6. Baking compatible -7. High temperature baking up to 250°C when the sensor unit is connected (when the sensor unit is disconnected: 300°C)
    8, DEGAS function-electron impact desorption method
    9, maintenance-easy replacement of filament, ion source and secondary electron multiplier
    10, enhanced preventive maintenance function-preventive maintenance of ion source and secondary electron multiplier, equipped with analysis tube traceability function (patent no 5016031)
    11, enhanced leak detection function-capable of helium leak test, air leak test and accumulation test
    12, capable of measuring total pressure
    13, standard equipped with Qulee QCS software
    14, conforming to CE

    | application

    1, thermal desorption analysis and other research and development
    2, ultra-high/ultra-high residual gas analysis of vacuum pumping equipment, etc.
    3. Measurement of trace impurities in gases
    4. Research and development related to solar cells
    5. Organic analysis
    6. Environmental measurement

  • | Specification Parameter

    Type HGM2-202 HGM2-302
    Mass Number Range 1-200 amu 1-300 amu
    Inspiration 2.5 × 10 -6 A/Pa,4A/Pa 2.0 × 10 -6 A/Pa,4A/Pa
    detects partial pressure 1 × 10 -9 Pa,1 × 10 -13 Pa 1 × 10 -9 Pa,1 × 10 -13 Pa
    using pressure 1 × 10 -2 Pa
    resolution m/m = 1m (10% p. h.)
    baking possible temperature
    (when sensor unit is connected)
    250 ℃
    Baking Possible Temperature
    (When Sensor Unit Removed)
    300 ℃
    Software Qulee QCS Ver. 4.0 to drop (Windows 7/8/10) apply